VICTOR Temperaure Calibrator
VICTOR 14+ Temperaure Calibrator
Features: High Accuracy of 0.02% for measure and source. 5 digits display for measure, and 6 digits display for
source. A dual, Liquid Crystal;Display White LED for backlight
Measuring Function
Function
|
Reference
|
Range
|
Resolution
|
Accuracy
|
Remark
|
DCV
|
50mV
|
-5.000mV~55.000mV
|
1μV
|
0.02+0.02
|
Input Resistan approximately 100MΩ
|
500mV
|
-50.00mV~550.00mV
|
10μV
|
0.02+0.01
|
OMH
|
500Ω
|
0.00Ω~550.00Ω
|
0.01Ω
|
0.05+0.02
|
Test current for 500Ω:Approximately
1mA;Test current for 5KΩ:
Approximately 0.1mA;Open circuit
voltage:About 2.5V;Dose not include
lead resistance;
|
5KΩ
|
0.0000 KΩ~5.5000KΩ
|
0.1Ω
|
0.05+0.02
|
TC
|
R
|
0℃~1767℃
|
1℃
|
0~500℃ : 1.8℃
|
By using ITS-90 temperature
scale;
The accuracy does not include
the error of internal temperature
compensation caused by a
sensor;
|
S
|
0℃~1767℃
|
500~1767℃ : 1.5℃
|
K
|
-100.0℃~1372.0℃
|
0.1℃
|
-100.0~0.0℃ : 1.2℃
|
E
|
-50.0℃~850.0℃
|
0.0~1372.0℃ : 0.8℃
|
J
|
-60.0℃~1120.0℃
|
-50.0℃~0.0℃ : 0.9℃
|
T
|
-100.0℃~400.0℃
|
0.0℃~85.0℃ : 1.5℃
|
N
|
-200.0℃~1300.0℃
|
-60.0~0.0℃ : 1.0℃
|
B
|
600℃~1820℃
|
1℃
|
0.0~1120.0℃ : 0.7℃
|
L
|
-60.0℃~900.0℃
|
0.1℃
|
-100.0~0.0℃ : 1.0℃
|
U
|
-100.0℃~600.0℃
|
0.1℃
|
0.0~400.0℃ : 0.7℃
|
RTD
CONTINUITY
|
PT100
385
|
-200.0℃~800.0℃
|
0.1℃
|
-200.0~0.0℃ : 1.5℃
|
By using temperature scale ITS-90.
Does not include lead resistance. Assuming all three RTD leads are matched for 3-w input.
|
RTD
CONTINUITY
|
PT1000
385
|
-200.0℃~630.0℃
|
0.0~1300.0℃ : 0.9℃
|
PT200
385
|
-200.0℃~630.0℃
|
600~800℃ : 2.2℃
|
PT500
385
|
-200.0℃~630.0℃
|
800~1000℃ : 1.8℃
|
Cu10
|
-100.0℃~260.0℃
|
1000~1820℃: 1.4℃
|
Cu50
|
-50.0℃~150.0℃
|
|
|
500Ω
|
≤50Ω sound
|
0.01Ω
|
|
500Ω Approximately 1mA test Current
|
TECHNICAL DATA (Output Function)
Function
|
Reference
|
Range
|
Resolution
|
Accuracy
|
Remark
|
DC voltage |
100mV |
-10.000~110.000mV |
1μV |
0.02+0.01 |
Maximum output: 0.5mA
Maximum output: 2mA |
1000mV |
-100.00~1100.00 mV |
10μV |
0.02+0.01 |
Resistance |
400Ω |
0.00Ω~400.00Ω |
0.01Ω |
0.02+0.02 |
Excitation current: ±0.05 –0.3mA;if ±0.1-0.5mA.add0.1Ω;
Does not include lead resistance; |
4KΩ |
0.0000 KΩ~4.0000 KΩ |
0.1Ω |
0.05+0.025 |
Excitation current: ±0.05 –0.3mA;
Does not include lead resistance; |
TC |
R |
0℃~1767℃ |
1℃ |
0~100℃: 1.5℃
100~1767℃: 1.2℃ |
By using ITS-90
temperature scale;The accuracy
does not include the error
of internal temperature
compensation caused by a
sensor; |
S |
0℃~1767℃ |
0~100℃: 1.5℃ |
100~1767℃: 1.2℃ |
K |
-200.0℃~1372.0℃ |
0.1℃ |
-200.0~-100.0℃:0.6℃ |
-100.0~400.0℃: 0.5℃ |
400.0~1200.0℃: 0.7℃ |
1200.0~1372.0℃: 0.9℃ |
E |
-200.0℃~1000.0℃ |
-200.0~-100.0℃: 0.6℃ |
-100.0~600.0℃: 0.5℃ |
600.0~1000.0℃: 0.4℃ |
J |
-200.0℃~1200.0℃ |
-200.0~-100.0℃: 0.6℃ |
-100.0~800.0℃: 0.5℃ |
800.0~1200.0℃: 0.7℃ |
T |
-250.0℃~400.0℃ |
-250.0~400.0℃: 0.6℃ |
N |
-200.0℃~1300.0℃ |
-200.0~-100.0℃: 1.0℃ |
-100.0~900.0℃: 0.7℃ |
900.0~1300.0℃: 0.8℃ |
B |
600℃~1820℃ |
1℃ |
600~800℃: 1.5℃ |
800~1820℃: 1.1℃ |
L |
-200.0℃~900.0℃ |
0.1℃ |
-200.0~0.0℃: 0.7℃ |
0.0~900.0℃: 0.5℃ |
U |
-200.0℃~600.0℃ |
0.1℃ |
-200.0~0.0℃: 0.7℃ |
0.0~600.0℃: 0.5℃ |
RTD |
PT100
385 |
-200.0℃~800.0℃ |
-200.0~0.0℃: 0.3℃ |
By using temperature scale
ITS-90
Excitation current: ±0.5~±3mA
for Pt100, Cu10, Cu50, add
0.5℃ when excitation current
is ±0.1mA-0.5mA;
Excitation current: ±0.05mA~
±0.3mA for PT200, PT500,
PT1000;
Does not include lead resistance |
0.0~400.0℃: 0.5℃ |
400.0~800.0℃: 0.8℃ |
PT1000 |
-200.0℃~630.0℃ |
-200.0~100.0℃: 0.2℃ |
100.0~300.0℃: 0.5℃ |
300.0~630.0℃: 0.7℃ |
PT200
385 |
-200.0℃~630.0℃ |
-200.0~100.0℃: 0.8℃ |
100.0~300.0℃: 0.9℃ |
300.0~630.0℃: 1.0℃ |
PT500
385 |
-200.0℃~630.0℃ |
-200.0~100.0℃: 0.4℃ |
100.0~300.0℃: 0.5℃ |
300.0~630.0℃: 0.7℃ |
Cu10 |
-10.0℃~260.0℃ |
-10.0~260.0℃: 1.8℃ |
Cu50 |
-50.0℃~150.0℃ |
-50.0~150.0℃: 0.6℃ |
Other feature:
1、Temperature Coefficient: 0.1 times the applicable accuracy specification per degree C
for 5℃to 18℃ and 28℃to 40℃.
2、 The range of the internal temperature compensation sensor is from 0℃to 50℃,
Compensation error ≤±0.5℃
3、 The accuracy of the temperature probe: ±0.2℃. The range of the measured temperature
is form -20℃-100℃.
4、Maximum voltage between any output terminal and earth: 30Vp-p