Toate Produsele
Temperaure Calibrator VICTOR 14+ VICTOR

VICTOR Temperaure Calibrator

Pret fara TVA 1 501,20 €
1、Temperature Coefficient: 0.1 times the applicable accuracy specification per degree C
for 5℃to 18℃ and 28℃to 40℃.
2、 The range of the internal temperature compensation sensor is from 0℃to 50℃,
Compensation error ≤±0.5℃
3、 The accuracy of the temperature probe: ±0.2℃. The range of the measured temperature
is form -20℃-100℃.
Stoc
-1
Cod
VICTOR 14+
Cantitate
  

VICTOR Temperaure Calibrator

 

                  VICTOR 14+  Temperaure Calibrator

Features: High Accuracy of 0.02% for measure and source. 5 digits display for measure, and 6 digits display for
source. A dual, Liquid Crystal;Display White LED for backlight

Measuring Function

Function

Reference

Range

Resolution

Accuracy

Remark

 

DCV

50mV

-5.000mV~55.000mV

1μV

0.02+0.02

Input Resistan approximately 100MΩ

500mV

-50.00mV~550.00mV

10μV

0.02+0.01

 

 

OMH

500Ω

0.00Ω~550.00Ω

0.01Ω

0.05+0.02

Test current for 500Ω:Approximately
1mA;Test current for 5KΩ:
Approximately 0.1mA;Open circuit
voltage:About 2.5V;Dose not include
lead resistance;

5KΩ

0.0000 KΩ~5.5000KΩ

0.1Ω

0.05+0.02

 

 

 

 

TC

 

R

0℃~1767℃

 

1℃

0~500℃ : 1.8℃

 

 

 

 

 

 

 

By using ITS-90 temperature
scale;
The accuracy does not include
the error of internal temperature
compensation caused by a
sensor;

S

0℃~1767℃

500~1767℃ : 1.5℃

K

-100.0℃~1372.0℃

 

 

 

0.1℃

-100.0~0.0℃ : 1.2℃

E

-50.0℃~850.0℃

0.0~1372.0℃ : 0.8℃

J

-60.0℃~1120.0℃

-50.0℃~0.0℃ : 0.9℃

T

-100.0℃~400.0℃

0.0℃~85.0℃ : 1.5℃

N

-200.0℃~1300.0℃

-60.0~0.0℃ : 1.0℃

B

600℃~1820℃

1℃

0.0~1120.0℃ : 0.7℃

L

-60.0℃~900.0℃

0.1℃

-100.0~0.0℃ : 1.0℃

U

-100.0℃~600.0℃

0.1℃

0.0~400.0℃ : 0.7℃

 

RTD
CONTINUITY

PT100
385

-200.0℃~800.0℃

 

 

 

0.1℃

-200.0~0.0℃ : 1.5℃

 

 

 

By using temperature scale ITS-90.
Does not include lead resistance. Assuming all three RTD leads are matched for 3-w input.

 

 

 

RTD
CONTINUITY

PT1000
385

-200.0℃~630.0℃

0.0~1300.0℃ : 0.9℃

PT200
385

-200.0℃~630.0℃

600~800℃ : 2.2℃

PT500
385

-200.0℃~630.0℃

800~1000℃ : 1.8℃

Cu10

-100.0℃~260.0℃

1000~1820℃: 1.4℃

Cu50

-50.0℃~150.0℃

 

 

500Ω

≤50Ω sound

0.01Ω

 

500Ω Approximately 1mA test Current

 

TECHNICAL DATA (Output Function)

Function

Reference

Range

Resolution

Accuracy

Remark

DC voltage 100mV -10.000~110.000mV 1μV 0.02+0.01 Maximum output: 0.5mA
Maximum output: 2mA
1000mV -100.00~1100.00 mV 10μV 0.02+0.01
Resistance 400Ω 0.00Ω~400.00Ω 0.01Ω 0.02+0.02 Excitation current: ±0.05 –0.3mA;if ±0.1-0.5mA.add0.1Ω;
Does not include lead resistance;
4KΩ 0.0000 KΩ~4.0000 KΩ 0.1Ω 0.05+0.025 Excitation current: ±0.05 –0.3mA;
Does not include lead resistance;
TC R 0℃~1767℃ 1℃ 0~100℃: 1.5℃
100~1767℃: 1.2℃
By using ITS-90
temperature scale;The accuracy
does not include the error
of internal temperature
compensation caused by a
sensor;
S 0℃~1767℃ 0~100℃: 1.5℃
100~1767℃: 1.2℃
K -200.0℃~1372.0℃ 0.1℃ -200.0~-100.0℃:0.6℃
-100.0~400.0℃: 0.5℃
400.0~1200.0℃: 0.7℃
1200.0~1372.0℃: 0.9℃
E -200.0℃~1000.0℃ -200.0~-100.0℃: 0.6℃
-100.0~600.0℃: 0.5℃
600.0~1000.0℃: 0.4℃
J -200.0℃~1200.0℃ -200.0~-100.0℃: 0.6℃
-100.0~800.0℃: 0.5℃
800.0~1200.0℃: 0.7℃
T -250.0℃~400.0℃ -250.0~400.0℃: 0.6℃
N -200.0℃~1300.0℃ -200.0~-100.0℃: 1.0℃
-100.0~900.0℃: 0.7℃
900.0~1300.0℃: 0.8℃
B 600℃~1820℃ 1℃ 600~800℃: 1.5℃
800~1820℃: 1.1℃
L -200.0℃~900.0℃ 0.1℃ -200.0~0.0℃: 0.7℃
0.0~900.0℃: 0.5℃
U -200.0℃~600.0℃ 0.1℃ -200.0~0.0℃: 0.7℃
0.0~600.0℃: 0.5℃
RTD PT100
385
-200.0℃~800.0℃ -200.0~0.0℃: 0.3℃ By using temperature scale
ITS-90
Excitation current: ±0.5~±3mA
for Pt100, Cu10, Cu50, add
0.5℃ when excitation current
is ±0.1mA-0.5mA;
Excitation current: ±0.05mA~
±0.3mA for PT200, PT500,
PT1000;
Does not include lead resistance
0.0~400.0℃: 0.5℃
400.0~800.0℃: 0.8℃
PT1000 -200.0℃~630.0℃ -200.0~100.0℃: 0.2℃
100.0~300.0℃: 0.5℃
300.0~630.0℃: 0.7℃
PT200
385
-200.0℃~630.0℃ -200.0~100.0℃: 0.8℃
100.0~300.0℃: 0.9℃
300.0~630.0℃: 1.0℃
PT500
385
-200.0℃~630.0℃ -200.0~100.0℃: 0.4℃
100.0~300.0℃: 0.5℃
300.0~630.0℃: 0.7℃
Cu10 -10.0℃~260.0℃ -10.0~260.0℃: 1.8℃
Cu50 -50.0℃~150.0℃ -50.0~150.0℃: 0.6℃

 

Other feature:

1、Temperature Coefficient: 0.1 times the applicable accuracy specification per degree C
for 5℃to 18℃ and 28℃to 40℃.
2、 The range of the internal temperature compensation sensor is from 0℃to 50℃,
Compensation error ≤±0.5℃
3、 The accuracy of the temperature probe: ±0.2℃. The range of the measured temperature
is form -20℃-100℃.
4、Maximum voltage between any output terminal and earth: 30Vp-p