VICTOR Process Calibrator
VICTOR 11+ - Process Calibrator
Features; High Accuracy of 0.02% for source, 6 digits display for source.A dual,
Liquid Crystal Display;White LED for backlight
Function |
Reference |
Range |
Resolution |
Accuracy |
Remark |
Thermocouple |
R |
0℃~1767℃ |
1℃ |
0~100℃:1.5℃ |
By using ITS-90 temperature scale;The accuracy does not include the error of internal temperature compensation caused by a sensor; |
100~1767℃:1.2℃ |
S |
0℃~1767℃ |
1℃ |
0~100℃:1.5℃ |
100~1767℃:1.2℃ |
B |
600℃~1820℃ |
1℃ |
600~800℃:1.5℃ |
800~1820℃:1.1℃ |
E |
-200.0℃~1000.0℃ |
0.1℃ |
-200.0~-100.0℃:0.6℃ |
-100.0~600.0℃: 0.5℃ |
600.0~1000.0℃: 0.4℃ |
K |
-200.0℃~1372.0℃ |
0.1℃ |
-200.0~-100.0℃: 0.6℃ |
-100.0~400.0℃: 0.5℃ |
400.0~1200.0℃: 0.7℃ |
1200.0~1372.0℃: 0.9℃ |
J |
-200.0℃~1200.0℃ |
0.1℃ |
-200.0~-100.0℃: 0.6℃ |
-100.0~800.0℃: 0.5℃ |
800.0~1200.0℃: 0.7℃ |
T |
-250.0℃~400.0℃ |
0.1℃ |
-250.0~400.0℃:0.6℃ |
N |
-200.0℃~1300.0℃ |
0.1℃ |
-200.0~-100.0℃: 1.0℃ |
-100.0~900.0℃: 0.7℃ |
900.0~1300.0℃: 0.8℃ |
L |
-200.0℃~900.0℃ |
0.1℃ |
-200.0~0.0℃:0.7℃ |
0.0~900.0℃:0.5℃ |
U |
-200.0℃~600.0℃ |
0.1℃ |
-200.0~0.0℃:0.7℃ |
0.0~600.0℃: 0.5℃ |
Other feature:
1、Temperature Coefficient: 0.1 times the applicable accuracy,specification per degree
C for 5℃to 18℃and 28℃to 50℃.
2、The range of the internal temperature compensation sensor is
from 0℃to 40℃, Compensation error ≤±0.5℃.
3、The accuracy of the temperature probe: ±0.2℃.
The range of the measured temperature is form -20℃-100℃.
4、Maximum voltage between any output terminal and earth: 30Vp-p
5、Maximum output current: Approximately 25mA.