All Products
Thermocouple Calibrator VICTOR 02 VICTOR

VICTOR Thermocouple Calibrator

From 491,40 €
Voltage 100mV, Measuring value ±30uV, Input Resistance 1MΩ, Temperature coefficient:±0.005%,range/°C(0°C-18°C,28°C-50°C)
Stock
-1
Code
VICTOR 02
Quantity
  

VICTOR Thermocouple Calibrator

                            

                         VICTOR 02 - Thermocouple Calibrator

 

Output Range Output range Resolution Accuracy  
Voltage 100mV -10.00 ~110.00mV 0.01mV ±0.05% Setting value +30uV The max. output current ±5mA
1000mV -100.00~1100.0mV 0.1mV ±0.05% Setting value ±0.3mV
Thermocouple R -40~1760℃ 1℃ ±0.05% Setting value ±3℃ (≤100℃)  Adopt the ITS-90 temperature 
S -20~1760℃ 1℃ ±0.05% Setting value ±2℃ (>100℃) standard (Note 1)
B 400~1800℃ 1℃ ±0.05% Setting value ±3℃ (400~600℃)    
±0.05% Setting value ±2℃ (>600℃)  
E -200.0~1000.0℃ 0.1℃ ±0.05% Setting value    
±2℃ (≤-100℃)  
K -200.0~1370℃ 0.1℃    
J -200.0~1200.0℃ 0.1℃ ±0.05% Setting value  
T -200.0~400.0℃ 0.1℃    
N -200.0~1300.0℃ 0.1℃ ±1℃ (>-100℃)  



Output Range Output range Resolution Accuracy
Voltage 100mV -10.00 ~110.00mV 10uV ±0.05% Measuring value ±30uV Input Resistance 1MΩ
Thermocouple R -40~1760℃ 1℃ ±0.05% Measuring value ±3℃ (≤100℃)  xInput tesistance1MΩ x 
S -20~1760℃ 1℃ ±0.05% Measuring value ±2℃ (>100℃) Adopt the ITS-90
B 400~1800℃ 1℃ ±0.05% Measuring value ±3℃ (400~600℃)  
 ±0.05% Setting value ±2℃ (>600℃) temperature standard  
E -200.0~1000.0℃ 0.1℃ ±0.05% Measuring value (Note 3)
 ±2℃ (≤-100℃)   
K -200.0~1370℃ 0.1℃    
J -200.0~1200.0℃ 0.1℃ ±0.05% Measuring value  
T -200.0~400.0℃ 0.1℃    
N -200.0~1300.0℃ 0.1℃ ±1℃ (>-100℃)  

 

Note:  

1.Not including the accuracy of inner temperature compensatory transducer.

The range of inner temperature compensatory transducer:-10-50°C,compensatory error 0.5°C.

2.Temperature coefficient:±0.005%range/°C(0°C-18°C,28°C-50°C)