All Products
Thermo Resistance Calibrator VICTOR 03 VICTOR

VICTOR Thermo Resistance Calibrator

From 491,40 €
Resistance 400Ω, Setting value ±0.2Ω, Measurong current 1mA, Power 9V battery. The range of exciting current:0.5mA~2mA,the max.output voltage≤2V
Stock
-1
Code
VICTOR 03
Quantity

VICTOR Thermo Resistance Calibrator

 

                    VICTOR 03  Thermo Resistance Calibrator

Input Range Input range Resolution Accuracy Explanation
Resistance 400Ω 0.0~450.0Ω 0.1Ω ±0.05% Setting value ±0.2Ω Measurong current 1mA Note1
Thermoresis-tance Pt100 -200.0~850.0℃ 0.1℃ ±0.05% Setting value ±0.6℃ 1mA exciting current use Pt100-385
Measuring current 1mA Note1.2
Cu50 -50.0~150.0℃ 0.1℃ 1mA excitong current Note1.2

 

Input Range Input range Resolution Accuracy Explanation
Resistance 400Ω 0.0~400.0Ω 0.1Ω ±0.05% Setting value ±0.2Ω ±1mA exciting current Note1.2
Thermoresis-tance Pt100 -200.0~850.0℃ 0.1℃ ±0.05% Setting value ±0.6℃ ±1mA exciting current use Pt100-85
Note1.2
Cu50 -50.0~150.0℃ 0.1℃ ±1mA excitong current Note1.2

 

Power 9V battery (ANSI/NEDA 1640A or IEC 6LR619 alkaline) or AC power adapter
Battery Approx. 20 hours under the condition of 10mA
Max.allowable voltage 30V
Operation temperture range 0-50°C
Operation humidity range ≤80%RH
Store temperature range ≤-10°C-50°C
Store humidity range ≤90%RH
Size 200×100×40mm
Weight 550g(with holster)
Accessories Operation manual,test line CF-36(probe with alligator clip)
Options AC adapter(VCPS),test line CF-31-A(probe clip)
Security Conforms to IEC 1010

 

Note:

1.Not including the accessorial lead resistance

2.The range of exciting current:0.5mA~2mA,the max.output voltage2V

3.Temperature coefficient:±0.005%range/°C(0°C-18°C,28°C-50°C)